Loading

FAQs

Get the most out of your enzyme with our documents, guides, brochures and tools

FAQs

Browse our frequently asked technical questions by selecting a product below.

General troubleshooting for low product or late Ct values

If you are observing unusually late Ct values, try diluting the template RNA. By doing this, you are diluting any inhibitors which may be present to a concentration where they do not inhibit the reaction. Additionally, try increasing the reverse transcription step to 55 °C and increasing the annealing/extension temperatures. This may help resolve difficulties caused by secondary structures present in the RNA template and/or primers.

In cases where reaction inhibition may be involved, try reducing the amount of template1 or add 0.4 – 4.4 mg/ml BSA to the reaction2.

For more specific problems contact [email protected] with the following information:

  • Amplicon size
  • Reaction setup
  • Cycling conditions
  • Screen grabs of amplification traces and melting profile

1  Scipioni et al. A SYBR Green RT-PCR assay in single tube to detect human and bovine noroviruses and control for inhibition. Virology Journal.5:94 (2008). doi: 1186/1743-422X-5-94

2  Plante et al. The use of bovine serum albumin to improve the RT‐qPCR detection of foodborne viruses rinsed from vegetable surfaces. Applied Microbiology. 52:3 (2010) doi: https://doi.org/10.1111/j.1472-765X.2010.02989.x