qPCRBIO SyGreen Mix FAQs

Can products generated with qPCRBIO SyGreen Mixes be digested, cloned, and sequenced?

Yes, PCR products generated with qPCRBIO SyGreen Mixes have the same characteristics as PCR products generated with wild-type Taq polymerase. They may be sequenced or digested with restriction endonucleases using standard protocols. Products are 3′-d(A)-tailed and may be used for TA cloning or may be blunt-ended or digested with restriction enzymes prior to cloning. For best results, we recommend purifying the PCR products using any standard PCR clean-up kit.

Do qPCRBIO SyGreen Mixes work with bisulphate converted gDNA template?

Yes. We have customers who have reported successful use of qPCRBIO SyGreen Mix on bisulphate converted gDNA.

Is the storage of sample DNA in 1x TE (10 mM Tris-HCl / 1 mM EDTA) buffer compatible with subsequent qPCR using qPCRBIO SyGreen mixes?

Yes, this storage buffer is compatible. The EDTA will chelate some of the magnesium in the mix, but not significantly enough to affect the reaction.

qPCRBIO SyGreen Mixes contain the dye SyGreen. Does it belong to the SYBR green family?

SyGreen belongs to the family of asymmetric cyanine dyes, which is the same family that SYBR™ Green belongs to. SyGreen is an improved, new generation dye from this family. Similar to SYBR™ Green, SyGreen intercalates the DNA. Unlike SYBR™ Green, SyGreen supports PCR amplification over a wide range of dye concentrations, that is, the dye does not inhibit the PCR reaction. This plays a role in producing robust and consistent DNA melting curves both with prokaryote and eukaryote templates, as it shows no preference in nucleotide composition.

What are the ROX concentrations in qPCRBIO SyGreen Mixes?

The qPCRBIO SyGreen Mixes that contain passive reference dyes come in different formulations, each with a different concentration of the passive reference dye:

  • qPCRBIO SyGreen Mix Lo-ROX (PB20.11) contains 112 nM ROX.
  • qPCRBIO SyGreen Mix Hi-ROX (PB20.12) contains 1.12 µM ROX.
  • qPCRBIO SyGreen Mix Fluorescein (PB20.13) contains 0.2 mM fluorescein.
  • qPCRBIO SyGreen Mix Separate-ROX (PB20.14) 2x mix contains no ROX and include a separate tube of 50 µM ROX additive. This enables you to choose what concentration of ROX you’d like to use.

You can use our qPCR Selection Tool under the Resources drop-down menu to determine which of our mixes are best suited for your qPCR machine.

What is included in qPCRBIO SyGreen Mixes?

qPCRBIO SyGreen Mixes are ready to use qPCR 2x Mastermixes. You only need to add primers, template DNA and PCR grade water during reaction set up.

What is ROX?

ROX is a passive reference dye which means it does not take part in the PCR reaction. It is used to normalise non-PCR related fluctuations in fluorescence.

What is the error rate of the HS Taq DNA Polymerase used in qPCRBIO SyGreen Mixes?

The enzyme has an error rate of approximately 1 error per 2.0 x 10⁵ nucleotides incorporated.

What is the MgCl2 concentration in qPCRBIO SyGreen Mixes?

All qPCRBIO SyGreen Mixes contain MgCl2 at a concentration of 6 mM. This means the final concentration in the reaction is 3 mM.

What should be considered if the normalisation of signal with ROX appears to be lower relative to a competitor’s mix?

Ensure you are using the right concentration of ROX because different instruments require different ROX concentrations. For example, if a qPCRBIO SyGreen Mix Hi-ROX is used with an instrument requiring a Lo-ROX mix, the software will normalise the qPCRBIO signal against the Hi-ROX level. This will significantly reduce the fluorescence level of the qPCRBIO mix relative to competitors’ mixes, if in that case the Lo-ROX mix was chosen for the competitor.

When comparing mixes from different manufacturers, it is better to carry out separate runs or turn the passive reference off before analysing data.

What troubleshooting is recommended if efficiency of amplification is reduced with standard dilutions?

It has been reported that efficiency can decrease with subsequent dilutions for the standard curve. We recommend avoiding this by diluting the standards in 10 mM Tris-HCl pH 8.0, 0.1 mM EDTA, 0.05% Tween-20. EDTA is a chelating agent and it plays a role in preventing DNAse activity1. Tween-20 is a detergent and prevents the DNA from adsorbing to the sides of the tubes2. Most microcentrifuges are made of polypropylene and research has demonstrated that DNA sticks very well to polypropylene3.

Standards should not be frozen after diluting them. Even in the presence of detergent, freezing seems to cause DNA to bind irreversibly to polypropylene. We suggest leaving your standards at 4°C and preparing a fresh batch every few weeks.

1  Barra, G. B. et al. EDTA-mediated inhibition of DNases protects circulating cell-free DNA from ex vivo degradation in blood samples. Clin Biochem 48, 976-981, doi:10.1016/j.clinbiochem.2015.02.014 (2015).

2  Linnarsson, S. Recent advances in DNA sequencing methods – general principles of sample preparation. Exp Cell Res 316, 1339-1343, doi:10.1016/j.yexcr.2010.02.036 (2010).

3  Gaillard, C. & Strauss, F. Avoiding adsorption of DNA to polypropylene tubes and denaturation of short DNA fragments. Technical Tips Online 3, 3 (1998).

What troubleshooting is recommended if the background signal is very high?

A high background level of fluorescence is most likely due to an excess of template in the reaction. This relates to the qPCR instrument picking up the SyGreen dye which binds to the template DNA. Samples usually contain a lot of DNA other than the target gene, so there will be enough present to register fluorescence. We recommend diluting the samples 100x-1000x to overcome this issue.

This should not only reduce the background signal, but also allow an accurate quantification because the dilution should bring the Ct values in the range of those obtained with the standards. Keep in mind that if the Ct is earlier than the first standard, it won’t be considered accurate enough quantification for publication.

Why is there a non-specific product when using the same primers and PCR conditions as a competitor’s product?

It’s most likely because the time for the 1st step (hot start) is too short. Ensure that the hot start phase is done at 95°C for 2 minutes to fully activate the enzyme. The recommended thermal profile is:

  • 95°C (120 seconds)
  • 40 cycles: 95°C (5-15 seconds) – 60°C (20-30 seconds)
  • Melt

If non-specific products are still obtained, we recommend raising the annealing/extension temperature from 60°C to 65°C, depending on the primer set used.

 

Why is there an earlier Ct when using qPCRBIO SyGreen Mixes compared to a competitor’s product?

An early Ct could indicate one of two things. The first could be the result of our product performing very well relative to the competitor’s product. We have designed our products to be faster than our competitors, and we have found that we are significantly faster than many.

The second could be that caused by non-specific products. This can be verified by running a melt analysis at the end of the qPCR run. If your results show one peak, it is likely the correct product was amplified. You may observe the position of the peak being different with our mix relative to our competitor’s mix but is this normal because the salt concentration and pH affect the melting temperature. If your results show more than one peak, then the reaction is not specific. PCR products can also be run on an agarose gel to verify if there is only one band, and if it is the same size as the amplicon obtained with the competitor’s product.

To address the issue of a non-specific product, the thermal cycling will need to be adjusted.